8 minutes · Public recording
Explore high-density EEG
Compare the same scalp map measured with every electrode of a 128-channel net, then with only 64 or 32 electrodes.
A high-density EEG net covers the whole head with electrodes, including the lower back of the head and the face. This recording comes from the Healthy Brain Network (HBN), a large open pediatric dataset from the Child Mind Institute. A participant watches discs appear on a screen. Each onset produces a visual response at the back of the head.
- Keep every electrode at 200 ms. Find the positive area at the back of the head. Switch to 32 electrodes. Does the overall shape stay the same? Read the error at the left-out electrodes.
- Turn on the surface Laplacian and compare again. Which map changes more when electrodes are removed?
- Pick an electrode and compare single trials with the average. Find it on the coverage diagram.
Compare scalp maps
- Reference map
Response strength over time
Standard deviation of the average across electrodes (global field power). Vertical lines mark the three map times.
Single trials and the average
Coverage
Seen from above, nose up. The colored dot is the electrode in the traces.
At a glance
- What it measures
- Voltage differences on the scalp, in microvolts. They come mostly from currents in large populations of pyramidal neurons in the cortex.
- Electrodes in this recording
- 128 electrodes plus the Cz reference, so 129 positions. The lab leaves out 25 noisy electrodes and keeps 104.
- Sampling rate
- 500 Hz as recorded. The lab shows the data at 125 Hz after a 20 Hz low-pass filter.
- What limits depth and resolution
- The skull conducts current poorly and spreads potentials across the scalp. Fine detail fades, and activity from a deep region reaches the scalp as a broad pattern.
- What full coverage buys
- Sampling dense enough to follow the map without gaps. Common estimates suggest 20 to 30 mm between electrodes, about what a 64 to 128 electrode system gives. Methods such as the surface Laplacian need close neighbors around each electrode.
Sources: NeuroTechX Primer, chapter 2 · Robinson and colleagues, 2017 · MNE surface Laplacian example
Where the maps come from
The lab uses both blocks of the surround suppression task from participant sub-NDARAG143ARJ. It keeps the 120 onsets where a stimulus is visible. The MNE-Python script filters from 0.5 to 20 Hz, which removes the discs’ 25 Hz flicker and keeps the response to their onset. It leaves out electrodes whose standard deviation is over three times the median in either block, applies an average reference, removes eye components found with ICA, cuts epochs from −200 to 600 ms and corrects the baseline. Autoreject keeps 116 trials.
The 64 and 32 electrode subsets start at Cz and then add the electrode farthest from those already chosen. Each map is interpolated with spherical splines onto the same grid. The Laplacian is computed again from each subset.
This is one recording from one participant. The electrode positions are those shipped with the files, not positions measured on this person’s head. The values illustrate the method in this recording.
Data by Shirazi, Franco, Scopel Hoffmann, Esper, Truong, Delorme, Milham and Makeig, HBN-EEG Release 1, Child Mind Institute. The excerpt in this lab is derived with MNE-Python and shared under the same CC BY-SA 4.0 license.
What density changes
In this recording, the potential map keeps its overall shape with 32 electrodes, but values between electrodes become less certain. The Laplacian changes more, because it measures differences between neighboring electrodes. The more fine detail a map has, the more electrodes it takes to follow it.